![horiba TPNA-500 Скачать руководство пользователя страница 232](http://html1.mh-extra.com/html/horiba/tpna-500/tpna-500_instruction-manual_2162279232.webp)
Troubleshooting
216
TP reading is inconsistent
TP reading is abnormal
The dilution tube or cell
measurement tube is dirty
Check if dirt is accumulated in the dilution
tube and cell measurement tube.
If they are dirty, clean the inside of the
glass tube.
“ Checking and
cleaning the
dilution tube and
cell
measurement
tube” (page 209)
Sample cannot be
measured
Check if the filter for the overflow tank is
dirty.
If it is dirty, clean it.
If the problem cannot be resolved, contact
HORIBA Advanced Techno.
“ Checking and
cleaning the
overflow tank
(option)” (page
207)
Cell measurement, blank
water measurement, or
reagent measurement
cannot be performed
appropriately
Perform each operation and report the
error condition to HORIBA Advanced
Techno.
-
The reagent has
deteriorated
Replace the reagent with a new one
designated by HORIBA Advanced
Techno.
“ Replacing the
reagent bottle”
(page 193)
The overflow tank is dirty
Clean the inside of the overflow tank and
the filter for the sampling port.
“ Checking and
cleaning the
overflow tank
(option)” (page
207)
Connector is disconnected
Check the connection of each connector.
If any of them is disconnected, connect it.
-
Symptom
Possible cause
Corrective action
Reference
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......