Preparation for Operation
31
Installation of reagent bottle
Note
Take care not to damage the reagent tray (load capacity: 5.0 kg)
While the reagent tray is pulled out, do not place anything other than the reagent bottle specified
by HORIBA Advanced Techno.
When setting or replacing the reagent, match the color of the tube connected to the nozzle unit
with that of the reagent bottle label. Then, set it in the correct location.
If it is set in an incorrect location, measurement cannot be performed correctly.
Reagent bottle
A label with different color is attached to each reagent bottle.
WARNING
Chemical hazard
Some reagents are poisonous and thus it is extremely dangerous if they are touched by hand. When
replacing the reagent bottle, be sure to wear protective goggles and gloves, and be extremely careful
so that your skin will not come in direct contact with reagents. When replacing the reagent bottle, the
reagent may spill off from the bottle. If it spills off, wipe it off completely.
Reagent
Reagent
capacity
Label color
Reagent A
1000 mL
White
Reagent B
250 mL
Blue
Reagent C
250 mL
Black
Reagent D
250 mL
Yellow
Reagent E
250 mL
Orange
Reagent F
250 mL
Red
Span
solution
1000 mL
Green