Material
235
Sequence: Zero calibration
Tip
A part of the display of “Drain Sample” in serial No. 1 varies depending on the setting conditions. For
details, refer to “ Drain Sample” (page 241).
Serial
No.
ZERO Cal.
Process
time (sec)
Accumulated
time (sec)
Accumulated
time
(min:sec)
Serial
No.
ZERO Cal.
Process
time (sec)
Accumulated
time (sec)
Accumulated
time
(min:sec)
1
Drain Sample-Chk
20
20
00:20
46
Stir P-cell
5
1006
16:46
2
Drain Blank
20
40
00:40
47
Standby
20
1026
17:06
3
Drain water P-cell
10
50
00:50
48
Standby
80
1106
18:26
4
Inject to P-cell
20
70
01:10
49
Stir N-cell
5
1111
18:31
5
Drain water P-cell
10
80
01:20
50
Waiting reaction
300
1411
23:31
6
Standby
26
106
01:46
51
Stir N-cell
5
1416
23:36
7
Gauge Blank Water-12
50
156
02:36
52
Waiting reaction
300
1716
28:36
8
Gauge Meas.Tube-2
5
161
02:41
53
Stir N-cell
5
1721
28:41
9
P Stir Meas.Tube
5
166
02:46
54
Waiting reaction
300
2021
33:41
10
Tube to P-cell-2
15
181
03:01
55
Stir N-cell
5
2026
33:46
11
Drain water P-cell
10
191
03:11
56
Line Blow L1
300
2326
38:46
12
Gauge Meas.Tube-2
5
196
03:16
57
Stir N-cell
5
2331
38:51
13
P Stir Meas.Tube
5
201
03:21
58
Waiting reaction
175
2506
41:46
14
Tube to P-cell-2
15
216
03:36
59
Standby
100
2606
43:26
15
Drain water P-cell
10
226
03:46
60
Gauge Blank Water-12
50
2656
44:16
16
Drain Blank
20
246
04:06
61
N Inject Reag. F
26.5
2683
44:42
17
Standby
80
326
05:26
62
Stir N-cell
5
2688
44:47
18
Standby
95
421
07:01
63
P Inject Reag. D
26.5
2714
45:14
19
Press Blow
3
424
07:04
64
Stir P-cell
5
2719
45:19
20
Gauge Meas.Tube1W
15
439
07:19
65
Cooling
130
2849
47:29
21
Tube to P-cell-1
15
454
07:34
66
P Measure Zero
40
2889
48:09
22
P Inject Reag. A
26.5
481
08:00
67
P Inject Reag. E
26.5
2916
48:35
23
P Inject Reag. C
26.5
507
08:27
68
Stir P-cell
5
2921
48:40
24
Stir P-cell
5
512
08:32
69
Degassing
260
3181
53:00
25
Drain Sample-Chk
20
532
08:52
70
N Measure Samp.
60
3241
54:00
26
Drain water N-cell
10
542
09:02
71
Wastewater N-cell
10
3251
54:10
27
Inject to N-cell
20
562
09:22
72
Gauge Meas.Tube-1
30
3281
54:40
28
Drain water N-cell
10
572
09:32
73
Tube to N-cell-1
15
3296
54:55
29
Standby
26
598
09:58
74
Wastewater N-cell
10
3306
55:05
30
Gauge Blank Water-12
50
648
10:48
75
Gauge Meas.Tube1W
15
3321
55:20
31
Gauge Meas.Tube-2
5
653
10:53
76
N Stir Meas.Tube
5
3326
55:25
32
N Stir Meas.Tube
5
658
10:58
77
Tube to N-cell-1
15
3341
55:40
33
Tube to N-cell-2
30
688
11:28
78
Drain water N-cell
10
3351
55:50
34
Drain water N-cell
10
698
11:38
79
Gauge Meas.Tube-2
5
3356
55:55
35
Gauge Meas.Tube-2
5
703
11:43
80
N Stir Meas.Tube
5
3361
56:00
36
N Stir Meas.Tube
5
708
11:48
81
Tube to N-cell-2
30
3391
56:30
37
Tube to N-cell-2
30
738
12:18
82
P Measure Samp.
50
3441
57:20
38
Drain water N-cell
10
748
12:28
83
Wastewater P-cell
10
3451
57:30
39
Drain Blank
20
768
12:48
84
Gauge Meas.Tube-1
30
3481
58:00
40
Standby
145
913
15:13
85
Tube to P-cell-1
15
3496
58:15
41
Gauge Meas.Tube1W
15
928
15:28
86
Wastewater P-cell
10
3506
58:25
42
Tube to N-cell-1
15
943
15:43
87
Gauge Meas.Tube-2
5
3511
58:30
43
N Inject Reag. A
26.5
970
16:09
88
P Stir Meas.Tube
5
3516
58:35
44
N Inject Reag. B
26.5
996
16:36
89
Tube to P-cell-2
15
3531
58:50
45
Stir N-cell
5
1001
16:41