Connection
19
Contact output
This equipment is equipped with the contact outputs described below.
Each output item is enabled when the relevant contact is set, and the following operation also
becomes available.
Output item
Standard specification
Item
Description
SYNCHRONOUS IDLE 1
Specified synchronous signal 1 is output.
SYNCHRONOUS IDLE 2
Specified synchronous signal 2 is output.
Calibration
Calibration is in progress.
Standby
System is standing by for measurement.
Cleaning
Cleaning is in progress.
Calib/Clean
Either calibration or cleaning is in progress.
Maintenance
Maintenance switch has been pressed. Although measurements continue
during maintenance, output is in accordance with each setting. Measurement
results are not compiled in the daily report.
WarmUp
Equipment warm-up in progress.
WarmUp-UV
UV meter warm-up in progress.
Power supply OFF
Power is not supplied.
TOTAL ALARM 1 to 6
A group of individually specified alarms 1 to 6 has been activated.
Int.memory Err.
An error has occurred in the memory inside the meter.
It indicates a batch of alarms including Factory Setting, User Setting, EEPROM
Read, CF Write, FROM Write, EEPROM Write, Measure History, Hour Report,
Day Report, Alarm History, Calib. History, and OperateData.
Clock Err.
An error has occurred in the clock of the meter.
Int. Power
An error has occurred in the power supply unit inside the meter.
Int.comm. Err.
A communication error has occurred in the meter.
Heat High Temp
The heater has overheated or is not operating.
Lamp Err.
An error has occurred in the light source inside the meter.
Temp. Sensor Err.
An error has occurred in the temperature sensor inside the meter.
Lack Blank
The pure water is insufficient.
Full Drain Tank
The tank has become full of waste liquid.
Lo Temp Machine
The usage environment is below the operating temperature range.
Heat Low Temp
The heater temperature is extremely low or the heater is not operating.
Lack Span
The span solution is insufficient.
Lack Reag.
The reagent is insufficient.
Cal. Err.
A calibration error has occurred.
Printer
An error has occurred in the printer inside the meter.
Water Filter
*
“Water Filter” and “Water Supply” are not supported.
Water Supply
Adc Read
An ADC reading error has occurred in the meter.
Press Error
Abnormal pressure due to leakage or clogging inside the equipment.
Measure L1
The L1 line is currently being measured.
Samp. Lack L1
Sample water is not being supplied through the L1 line.
TP Hi Conc. L1
The TP concentration for the L1 line has exceeded the upper limit value.