Calibration
55
Calibration procedure
Points to check before calibration
Check the following.
Sufficient reagent and pure water remain
The calibration cycle and count are set (refer to “ AUTO CAL. SET” (page 70))
Operation procedure
1.
Press [START] in the main screen.
2.
Zero calibration and span calibration will be performed automatically. When
calibration is completed, measurement will be started according to the settings.
[START]
[CALIBRATION]