Troubleshooting
222
Alarm 5
VIS SPAN Cal.
Incompatible span
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Use the dedicated span solution. Be
sure to use distilled water for the
dilution water.
Clean the measurement cell.
If the wiper has deteriorated, replace
it.
-
Alarm 5
WarmUp-PN
The analyzer is waiting
until it is stabilized after
the power has been
turned ON
It is not an error. The analyzer requires
at least 10 minutes to warm up.
-
Alarm 5
WarmUp-UV
The analyzer is waiting
until it is stabilized after
the power has been
turned ON
It is not an error. It takes 60 minutes
until the analyzer is stabilized.
-
Alarm 5
Maintenance
The maintenance switch
has been pressed
It is not a meter error. Turning the
maintenance switch OFF returns the
meter to normal operation.
Alarm 5
Maintenance
UV
The maintenance screen
for the UV meter is
displayed
It is not a meter error. Switching to
another screen from the maintenance
screen on the UV meter returns the
meter to normal operation.
Alarm 5
CF Write
The CF memory is
insufficient or the CF
memory is faulty
Contact HORIBA Advanced Techno.
-
Alarm 5
FROM Write
FROM writing has failed
Contact HORIBA Advanced Techno.
-
Alarm 5
EEPROM Write
EEPROM writing has
failed
Contact HORIBA Advanced Techno.
-
Alarm 5
Printer
Printer paper has run out,
cover is open, or
connection is faulty
If the printer runs out of paper,
replenish the printer paper.
If the cover is open, close it.
For any problems other than those
above, contact HORIBA Advanced
Techno.
“ Preparation of
printer” (page
30)
Alarm 5
Usb Detect
The USB memory device
cannot be detected.
Re-check that the USB memory device
is correctly inserted.
Alarm 5
Usb Full
The USB memory
capacity is insufficient
The USB memory is
faulty
Re-check that the USB memory device
is correctly inserted. Check that the
memory capacity is at least 500 MB.
-
Alarm 5
1Year Check
Day
Annual inspection is now
overdue.
Contact HORIBA Advanced Techno.
-
Alarm 5
3Year Check
Day
Three-year inspection is
now overdue.
Contact HORIBA Advanced Techno.
-
Alarm 5
5Year Check
Day
Five-year inspection is
now overdue.
Contact HORIBA Advanced Techno.
-
Alarm 5
Water Filter
The pure water cartridge
has deteriorated
Replace the pure water cartridge.
*4
Alarm 5
Water Supply
The solenoid valve for
pure water supply is
faulty
Pure water
measurement failure
Contact HORIBA Advanced Techno.
*5
Classification
Alarm
description
Possible cause
Corrective action
Reference
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......