Function
85
SYNCHRONOUS IDLE
Specifies the synchronous signal.
Tip
If you wish to output a signal before the sequence starts, specify a negative value. For example, when
10 sec is specified, the signal is output 10 seconds before the sequence starts.
Operation procedure
1.
Press [EXT. POINT OUTPUT SET] in the OUTPUT SET screen.
2.
Press [SYNCHRONOUS IDLE] in the EXT. PO SET screen.
The SYNCHRONOUS IDLE screen will be displayed.
3.
Specify the synchronous signal.
Operation mode
Setting
range
Default
value
Description
SYNCHRONOUS
IDLE 1
ON
-3,600 sec
to 3600 sec
0 sec
The time period for which the signal is turned ON in
synchronization with the measurement sequence.
OFF
3600 sec
The time period for which the signal is turned OFF in
synchronization with the measurement sequence.
SYNCHRONOUS
IDLE 2
ON
-3600 sec
to 3600 sec
0 sec
The time period for which the signal is turned ON in
synchronization with the measurement sequence.
OFF
3600 sec
The time period for which the signal is turned OFF in
synchronization with the measurement sequence.
Main
MENU
SETTING
OUTPUT SET
[MENU]
[OUTPUT SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
[EXT. POINT
OUTPUT SET
EXT. PO SET
[SYNCHRONOUS IDLE]
OUTPUT SET]
[SYNCHRONOUS
[SYNCHRONOUS
SYNCHRONOUS IDLE
SYNCHRONOUS IDLE 1
SYNCHRONOUS IDLE 2
ON/OFF
ON/OFF
<Input a value>
<Input a value>
IDLE 1]
IDLE 2]
[ ]
[ ]