Maintenance
207
Checking and cleaning the overflow tank (option)
Periodically check the overflow tank and clean it if it is dirty.
Note
If there is an excessive amount of suspended solids (SS) in the sample water, periodically clean
the tank before the suspended solids accumulate inside the tank.
When you perform maintenance, be sure to stop the measurement, and then turn the
maintenance switch ON.
Inspection
Check that the overflow water level in the water tank is constantly maintained at the specified
level (water flows into the overflow tube).
Target
Overflow tank
Procedure
Inspection/cleaning
Cycle/timing (guide)
Depends on usage conditions