Modbus
187
CONTROL (EXT. ANALOG OUTPUT CHECK)
1.
Setting of output value (values can be read)
2.
Output request: Output check (0) is set and then output.
3.
Output request: Output check completion (1) is set and then the output check
status is canceled.
Note
Output check status is valid for 10 minutes after the receipt of output check request (0).
CONTROL (EXT. POINT OUTPUT CHECK)
1.
Setting of output value (values can be read)
2.
Output request: Output check (0) is set and then output.
3.
Output request: Output check completion (1) is set and then the output check
status is canceled.
Note
Output check status is valid for 10 minutes after the receipt of output check request (0).
ADJUST (EXT. ANALOG INPUT ADJUST)
1.
Setting of CH
2.
Setting of adjustment mode
3.
Adjustment request: Data set (0) is set and then the current value is read.
4.
Adjustment request: Setting update (1) is set and then the setting is updated.
5.
Adjustment request: Completion (2) is set and then the adjustment status is
canceled.
Note
The input adjustment status is valid for 10 minutes after the receipt of the adjustment request.
Be sure to execute “Adjustment request: Data set (0)” before executing “Adjustment request:
Setting update (1)”.
History (Measure History)
1.
Setting of request start time
2.
Setting of request completion time
3.
Request status: The same data (3) are set and then the first data are read.
4.
Request status: The next data (2) are set and then the data are read.
Note
Request status: When the same data (3) are set, the same data are read.