Function
119
SEPARATE ACTION
Allows you to perform each action individually.
Screen
Operation name
Description
SEPARATE
ACTION 1/5
N Gage Reag. A
Measures the reagent A line (TN).
P Gage Reag. A
Measures the reagent A line (TP).
Gage Reag. B
Measures the reagent B line.
Gage Reag. C
Measures the reagent C line.
Gage Reag. D
Measures the reagent D line.
SEPARATE
ACTION 2/5
Gage Reag. E
Measures the reagent E line.
Gage Reag. F
Measures the reagent F line.
Gauge Span Water
Measures the span solution line.
Gauge Samp.L1
Measures sample water from line 1.
Wastewater P-cell
Drains the residual liquid from the P cell into the waste liquid tank.
SEPARATE
ACTION 3/5
Drain Reag. A
Drains the residual reagent from the reagent A line into the reagent A
bottle.
Drain Reag. B
Drains the residual reagent from the reagent B line into the reagent B
bottle.
Drain Reag. C
Drains the residual reagent from the reagent C line into the reagent C
bottle.
Drain Reag. D
Drains the residual reagent from the reagent D line into the reagent D
bottle.
Drain Reag. E
Drains the residual reagent from the reagent E line into the reagent E
bottle.
SEPARATE
ACTION 4/5
Drain Reag. F
Drains the residual reagent from the reagent F line into the reagent F
bottle.
Drain Span Water
Drains the residual reagent from the dilution tube or span solution line
into the span solution bottle.
Drain Blank
Drains the residual liquid from the pure water measurement tube or
blank water line into the pure water tank.
Drain Sample L1
Drains the residual liquid from the dilution tube or sample line L1 to the
sample port L1.
Wash cell-line
Cell measurement -> P cell -> waste liquid tank (repeated 3 times)
Cell measurement -> N cell -> waste liquid tank (repeated 3 times)
Carry out cleaning six times without interruption as above.
SEPARATE
ACTION 5/5
Pre-Wash
Performs the pre-wash operation. However, the waste liquid is drained
into the waste liquid tank.
Tip
“ Sequence: Preprocess, Pre-Wash, Post-Wash” (page 234)
Preprocess
Performs the pre-process operation.
Tip
“ Sequence: Preprocess, Pre-Wash, Post-Wash” (page 234)
Discharge all
Drains the residual liquid from all the reagent lines, sample water line,
blank water line, and measurement cell into the waste liquid tank.
Pure water purge
Provided for models equipped with the optional built-in pure water unit.
It purges the pure water cartridge.
Check Press
Allows you to check the pressure inside the meter.