Calibration Menu
UVTOCOperation.fm
Section 6
TOC mg/L. Three of the calibrations, the primary calibration, the manual
calibration, and the auto calibration are two point calibrations as described
above. Ideally they would be all be the same data points, however minor
variations will exist due to tolerances. These calibrations should also all yield
a result of the same gain (slope). The fourth calibration is a multiple point
calibration arrived at by measuring the CO
2
ppm produced by more than two
standards, one of them always being the zero.
Primary Calibration and Manual Calibration
Primary and Manual Calibration are identical supervised calibrations that are
run by the operator. The distinction between the two is that a primary
calibration should be performed only after the peristaltic or lamp has been
replaced, when the analyzer will have the best performance and theoretical
gain. A successful primary calibration will automatically update the manual
calibration data set as well as the primary calibration data set.
Auto-calibration
This scheduled calibration sequence is automatically carried out by the
analyzer. Zero and span standards are automatically drawn from the ports
as required, to determine the auto-calibration data set. A successful
auto-calibration will become the active calibration.
Multi-point calibrations
The three forms of multi-point calibrations are detailed in
Active calibration
This is the calibration data set used to calculate the TOC during an online
TOC measurement. The primary, manual, autocalibration or multipoint
calibration may be manually set to be the active calibration.
The successful completion of any type of calibration will result in the
calibration being used as the active calibration and the entry in the manual
control/active calibration screen will be changed to reflect the type of
calibration just completed.
Active calibration versus primary calibration % (Prim v cal)
This is the ratio of the gains of the active calibration set to the gain of the
primary calibration set expressed as a percentage. The deviation from 100%
will identify the change in analyzer gain.
If the active calibration set gain:
g
a
=
is compared to the primary gain
g
p
=
as a percentage,
the result is:
After a primary calibration, this percentage will always be 100% because the
active calibration is the primary calibration. If manual calibration is selected, it
will still be 100% as the primary calibration copies to the manual calibration
set. Only following the next calibration will the result be different. Thus a
y
2a
y
1a
–
(
)
x
2a
-------------------------------
y
2p
y
1p
–
(
)
x
2p
-------------------------------
100
g
a
•
g
p
----------------------
100
x
2p
•
y
2a
y
1a
–
(
)
•
x
2a
y
2p
y
1p
–
(
)
•
---------------------------------------------------------------
=
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...