Section 5
Page 83
UVTOCSetup and System Configuration.fm
Setup Menu
Register assignment for ModBus is in compliance with the required network
specification. More information about the ModBus Protocol can be found on
the Schneider Automation website at www.modicon.com. The Modicon
ModBus Protocol Reference Guide can also be ordered from Modicon using
part number PI-MBUS-300.
5.2.8 Tolerances Screen
The Tolerance menu allows setting up of analog quantities, which largely
determine analyzer performance. It is recommended that they not be changed
from their factory settings.
5.2.8.1 Zero Tolerance Screen
Zero tolerance is acceptable CO
2
ppm variation (+/-) allowable when an
autocalibration zero operation is being performed.
Press the Up button while in the Zero tolerances screen. The Span tolerance
screen is displayed.
Table 12 Modbus Registers
Register address
Function
40001/2
Live output in mg/L, single precision IEEE float
40003/4
Gas PPM output, single precision IEEE float
40005/6
Gas PPM slope in ppm/min, single precision IEEE float
40007
Reactor temperature in degrees Celsius, integer
40008
Reactor pressure in Kilo-Pascals, integer
40009
Currently selected stream, integer
40010
Machine status, integer
40011
Measurement sync counter (modulo 256), integer
40012
Epa status result, boolean
40013
RTC register 1, Year, YYYY
40014
RTC register 2, Month/day MMDD
40015
RTC register 3, Hour/minute HHMM
40016
RTC register 4, Seconds xxSS
40017
Retention time in minutes, integer
40018/19
Removal percent, single precision IEEE float
40020/21
Alkalinity setpoint, single precision IEEE float
40022/23
EPA TOC output 1, single precision IEEE float
T o l e r a n c e s
Z e r o t o l e r a n c e
5 p p m
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...