Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
5.2.6.1 Map Analog Outputs
Each of the two analog outputs can be mapped to one of the
following variables:
Range ID:
Range ID allows monitoring the analyzer's current analog output range.
This is necessary if autoranging is selected
Fault ID:
Fault ID allows the online, offline, maintenance, and fault status of the
analyzer to be monitored on the analog output. The following convention
is used:
For status codes, fault IDs, and fault
codes, refer to
.
Latched mg/L 1:
Latched mg/L measures carbon concentration and holds the last valid online
reading when the instrument is taken offline.
Zero mg/L 1:
Zero mg/L forces all output to zero when offline. Tracks TOC when online.
Gas ppm:
The CO
2
gas ppm may be output from zero to full scale, linearized, and the IR
cell temperature may be directed to the analog output.
Live mg/L:
Live mg/L measures carbon concentration, even during offline periods,
allowing the output to track the single sample, calibrations standards, etc.
Temperature:
The IR cell output may be tracked on the analog output (0 –100 °C)
Zero to Span:
Normally only used for fault finding
Slope:
The rate of change of mg/L/sec.
M a p a n a l o g s
Range 1
4 mA
Range 2
8 mA
Range 3
12 mA
Range 4
16 mA
Fault ID
Fault Description
4 mA
Online.
8 mA
Out of service (the analyzer is offline, purging, cleaning,
calibrating, or in the Service screens).
12 mA
Maintenance request, e.g., an attempt at calibration has failed.
16 mA
Fault.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...