Mirror Assembly Adjustment
Figure 50 UV Lamp Assembly with Cover Removed
7.12
Mirror Assembly Adjustment
Disconnect the power to the relays before opening the analyzer
enclosure doors and read
Electrostatic Discharge (ESD) Considerations on
1.
Locate the 8000 Display/Controller PCB assembly (mounted on the door
of the electronics cabinet). Remove the cover.
2.
Refer to
and locate test points TP3 (0V) and TP5 (SIG IN).
3.
Set up the instrument:
Using an Oscilloscope (recommended):
a.
Connect the signal probe to TP5 and the ground lead to TP3.
b. Set the display to 1 volt per division.
Using a Volt meter:
a.
Connect the positive lead to TP5 and the negative lead to TP3.
b. Set the volt meter to ac.
The ac signal at TP5 alternates
between 1 and 2 Hz and is
asymmetric. The asymmetry will
cause the volt meter reading to vary
continuously as much as 400 mV.
This is normal.
4.
Apply carrier gas to the IR inlet at approximately 100 mL per minute. Wait
2–5 minutes before checking the signal level.
5.
Using a nut driver, adjust the three hex nuts at the end of the IR cell for the
highest possible voltage level:
Oscilloscope: 5 to 7.5 volts peak–to–peak within 5% with each other
Volt meter: 1.5 to 2.3 volts
6.
When you have finished, calibrate the analyzer.
1.
UV Lamp (2)
2.
Lamp Inverter (2)
UP
1
2
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...