Section 6
Page 103
UVTOCOperation.fm
Calibration Menu
only reflect the calculation based on the most recent calibration point
measured and the multipoint zero.
Segmented
When this option is selected, the multi-point zero and the non zero multi-point
standards will be used to calculate the TOC using a piece point linear
representation. See
See Safety Note 32 on page 103
.
Figure 32 Segmented—Piece-point Linear Representation
TOC added
If the standards used for the TOC added calibration are analyzed after the
calibration online, the TOC reported will be offset by the reagent water blank.
If the standards used for the TOC
added calibration are analyzed after
the calibration online, the TOC
reported will be offset by the
reagent water blank.
When this option is selected, the multi-point zero is not used. Only the non
zero TOC calibration points are used to calculate the best-fit straight line
calibration. See
This line is used to calculate the reagent water blank and its associated ppm
CO
2
value. A TOC added calibration is useful when calibrating a low range
analyzer when the unknown carbon impurity in the DI water is significant.
Fitted
When this option is selected the multi-point zero and the multi-point
TOC standards will be used to calculate the TOC using the best-fit
straight-line calibration. See
.
TOC µg/L or mg/L
CO ppm
2
0
20%
20%
40%
60%
80%
100%
40%
60%
80%
100%
0
(0, y )
1
(x , y )
6
6
(x , y )
5
5
(x , y )
4
4
(x , y )
2
2
(x , y )
3
3
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...