Section 6
Page 101
UVTOCOperation.fm
Calibration Menu
The span value is not necessarily
the full scale span, but is the value
of the span calibration standard.
The displayed values (in ppm and
mg/L) may not match your analyzer.
2.
Press ENTER to progress to the prompt screen to ensure the standard is
connected to the calibration port.
3.
Press ENTER. The screen shows the apparent TOC using the prior
calibration, the pre span countdown timer, and the prior span TOC mg/L.
Press UP to display the equivalent screen displaying ppm CO
2
instead of
mg/L TOC.
4.
After the pre span countdown is completed, the analyzer displays the
TOC mg/L reading and prompts to press ENTER when stable.
For an accurate calibration, high range analyzers benefit from
long exposure to the span standard. The stability is tested before
determining the result, however, the user is responsible for determining
when the reading is stable. During this period, press ENTER to progress
to the next screen.
5.
Press ENTER to store the result of the calibration in the analyzer
data base. A comparison screen between the live ppm and the span
ppm results.
At this point the manual calibration is made the active calibration, i.e.,
Cal to use is set to Manual.
6.
Press ENTER to produce a final screen that displays the comparison of
the calibration set with the primary calibration data set.
A p p l y S T D t o
c a l i b r a t i o n p o r t
9 5 0 m g / l
126s
L a s t
1 0 0 0 m g / l
1 0 0 0 m g / l
STAB
L a s t
1 0 0 0 m g / l
L i v e
9 5 0 3 p p m
S p a n
9 5 0 1 p p m
1 0 0 0 m g / l
1 0 1 . 5 % P r i m .
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...