Setup Menu
5.2.1.3 Range Change % Screen
Use this function only when using the autoranging function. If you are using a
fixed range, disregard this setting.
The 4–20 mA analog outputs can be used as autoranging scaled outputs.
As the TOC level increases, that increase is translated into a percentage of
the user-defined range (see
). The outputs automatically
change to the next higher range at a defined percentage setting. The other
analog output should be set to reflect the Range ID.
If the TOC level is decreasing, the range will change at 5% lower than the
defined percentage. This avoids rapid range changes when the reading is
stationary and close to the Range change %.
Changes to higher and lower ranges are discussed in detail later in
this section.
1.
To enter the Range change % screen, press the UP button while in the
Inhibit level screen.
2.
To modify the Range change%, press ENTER.
3.
Asterisks will flash on either side of the displayed value. Use the UP and
DOWN buttons to change the value.
4.
Press ENTER when the desired value is displayed. The new value is
entered into memory.
For example: If you have a 0 to 1000 mg/L analyzer, the ranges could be
divided as follows:
If range change is set at 90%, the mapped analog output changes from
Range 1 to Range 2 at 90 mg/L (90% of 100 mg/L). The change from Range 3
to Range 4 occurs at 450 mg/L (90% of 500 mg/L).
When the measurement is descending, (Range 4 to Range 3, etc.),
the range change occurs at a point 5% below the programmed value. Using
the previous example, the change from Range 4 to Range 3 occurs at
425 mg/L (85% of 500 = 425). The change from Range 3 to Range 2 occurs
at 212 mg/L (85% of 250 = 212). The change from Range 2 to Range 1 occurs
at 85 mg/L (85% of 100 = 85).
R a n g e c h a n g e
%
85%
Range 1 starts at 0 mg/L and goes to 100 mg/L
Range 2 starts at 0 mg/L and goes to 250 mg/L
Range 3 starts at 0 mg/L and goes to 500 mg/L
Range 4 starts at 0 mg/L and goes to 1000 mg/L
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...