Section 3
Page 53
UVTOCReagents and Standards.fm
Storage and Handling Tips
Adjust the following as necessary:
1.
Interval, duration, and day of cleaning action. Refer to
.
2.
Perform cleaning with the UV-lamps turned ON or OFF. Refer to
below and
3.
Establish the cleaning solution.
When establishing the cleaning solution, it is important to start with the
least aggressive or abrasive liquid such as DI water. DI water will often be
adequate, but may require more frequent cleaning cycles. Other potential
cleaning solutions are:
•
50/50 mix (by volume) of H
3
PO
4
and Na
2
O
8
S
2
(UV-lamps ON or OFF)
•
1.0 M sodium hydroxide
•
5% NaOCl/CLOROX (UV-lamps OFF)
•
10% HCl (UV-lamps OFF)
•
5% Acetone (UV-lamps OFF)
If a cleaning agent has been used in another application containing the
same substances found in the sample(s), that cleaning agent can usually
be applied in this analyzer. Take caution when choosing the cleaning
solution. Take into consideration the materials that come in contact with
the sample and any gases or by-products produced.
3.8 Storage and Handling Tips
Even carefully prepared low TOC water can be easily contaminated by
atmospheric CO
2
. Follow the guidelines below to slow contamination:
•
Use glass containers for mixing and storing.
•
When not in use, seal calibration standards within an inert atmosphere so
that ambient CO
2
will not contaminate the solutions. Plastic membranes,
such as Parafilm
®
, will not keep out carbon dioxide.
•
Refrigerate samples containing volatile components.
•
Basic solutions absorb more ambient CO
2
than complex solutions.
•
Cold water absorbs more ambient CO
2
than hot water.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...