Page 122
IR Gas Calibration
UVTOCMaintenance.fm
Section 7
must be out. If the span gas is 10000 ppm CO
2
, LK1 must be in. See
Zero Adjustment
1.
The Zero adjustment is made on the 8000 controller PCB, which is
mounted inside the front door of the electronics cabinet.
illustrates the zero potentiometer adjustment. Turn the screw clockwise to
increase the voltage and counterclockwise to decrease the voltage.
After the zero gas reading has stabilized, adjust the Zero potentiometer
so that the voltage is between 0.26 and 0.27 V. 0.27 V will produce a CO
2
ppm reading slightly above zero. If the voltage is set at 0.25 V, a zero
reading would result, however you could not be certain if the gas ppm
reading was zero or below zero. The slight gas ppm increment above zero
ensures that you are in a positive range.
Figure 47 Adjusting the Zero Potentiometer on the 8000 Controller PCB
Span Adjustment
1.
Connect the appropriate span gas (1000 ppm or 10000 ppm) to the IR
and let it flow for ten minutes at 200 cc/minute until the reading stabilizes.
2.
After the reading stabilizes, adjust the span potentiometer to read the
approximate CO
2
ppm, either 1000 or 10000 ppm.
If the span gas does not yield 1000 ppm ± 5 ppm, or 10,000 ppm ± 50
ppm, verify the LK1 jumper position, see
, and if
necessary contact your manufacturer representative.
Noise will create minor fluctuations in the readings, however if these
fluctuations bracket the desired values, they are acceptable.
I R c a l
49 .9 C
1 7 p p m
0 . 2 6 V
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...