8.1 Preliminary Check
Before referring to the troubleshooting guide, make a routine check for
simple problems:
1.
Confirm proper electrical power to the analyzer.
2.
Check the fuses.
3.
Verify that liquid and gas supplies are present and adequate.
4.
Check all drains, sample, and gas lines to make sure that there are no
kinks, restrictions, or leaks.
If necessary, refer to the flow diagram,
.
8.2 Troubleshooting
DANGER
The tasks presented in this section of the manual may have hazards—including
electrical shock—associated with them. Because they require specialized
training and knowledge, they should only be performed by qualified personnel.
CAUTION
Remove power to the relays before performing tasks in this manual section.
Depending on how the relays are configured into your data communication
system, disconnecting power may cause a fault condition to be seen.
Important Note: Delicate internal electronic components can be damaged by static
electricity, resulting in degraded instrument performance or eventual failure.
8.2.1 Instrument Does Not Calibrate
1.
Check the flow by placing your finger over the IR cell outlet (vent) and
observing the GLS U–tube. Gas bubbles should circulate in the U–tube
from front to back.
Liquid tests requiring changes to the
plumbing should be done following
an analyzer flush to avoid contact
with samples and reagents.
2.
Ensure that the overflow portion of the GLS is filled with water and is
draining. The U-tube in the GLS may provide a simple indication of a clog
or obstruction in the line containing the condenser and IR measurement
cell. Visually observe that the water level in the back-pressure gauge on
the GLS is at the operational level. An increase in pressure (caused by
the clog or obstruction) will push the level of the U-tube liquid seal down
on the right side until gas bubbles around the bottom of the U–tube
appear. If bubbles are forced through the drain, there is excessive back
pressure in the IR or vent line. Inspect and clean or replace the IR tubing.
3.
In order to verify carrier flow integrity, the system should be flushed of all
sample and reagents for 15 minutes before continuing. Stop the analyzer.
Loosen the carrier flow supply on the UV manifold. Fold the carrier supply
tubing to obstruct flow or block with a fitting. The carrier flow should fall to
zero, as shown by the flow meter, and an alarm may be generated as the
reactor has been pressurized.
4.
Fit a plug into the sparger block manifold outlet going to the resample
pump. Place the inlet to the resample pump into a new calibration
standard. In dilution analyzers it will be necessary to dilute the standard
by the dilution factor, normally 10. Tubing from the start-up kit may be
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...