Page 136
Fault Code Diagnostics
UVTOCTroubleshooting and Service.fm
Section 8
Drain Level
The analyzer has detected that the level sensor
in the drain pan has been actuated.
—
The analyzer has a leak and the drain is
blocked.
Verify drain is clear.
The level sensor is installed with the wrong
orientation.
Invert level sensor.
F05, F07, and F08
Not Allocated.
F06
Reactor blocked
The pressure in the UV reactor has exceeded
the user specified fault pressure threshold
Setup/Tolerances.
The reactor is blocked. Locate and remove
blockage. Possible causes are:
1.
UV manifold crystal buildup in the UV
block. Clean the UV manifold.
2.
Blocked UV reactor. Disassemble the
reactor and flush with water. If the blockage
is not removed with flushing, contact
technical support.
F09
IR detection
The Automatic Gain Control level on the IR has
reached an unacceptable level
Inspect and clean optics in IR cell. See
F10-F16
Not allocated.
1
If this event persists, it may be disabled in the Service/Input Status Screen. See
Code
Fault Message
Diagnosis
Corrective Action
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...