EPA Mode
UVTOCIntroduction.fm
Section 1
1.10 EPA Mode
The selection of the EPA application allows the analyzer to be operated
compliant with the requirements of the EPA regulation (Part IV- Environmental
Protection Agency/40 CFR Parts 9, 141 & 142). TOC
1
is TOC concentration in
mg/L of the incoming raw water source. TOC
2
is the TOC concentration in
mg/L of the processed (distributed) water. The removal percentage (REM%)
is calculated as follows:
Where TOC
1
is measured a period of at least the retention time earlier than
the measurement of TOC
2
. The TOC data points are retained at the time that
the analyzer goes offline. Only the last Stream 2 TOC value is recorded for the
REM% calculation, however the analyzer records a history of the Stream 1
TOC values.
The pass or fail criteria is based on the manually entered incoming water
alkalinity value and the analyzed incoming source water TOC concentration
(TOC
1
). A pass/fail can be determined and shown on the normal operating
screen according to the value of the required removal percentage. The
following table details how the pass/fail criteria is established from the entered
alkalinity (as mg/L C) and the two TOC readings.
REM%
100
=
TOC
1
TOC
2
–
(
)
TOC
1
-------------------------------------------
×
Table 1 Establishing the Pass/Fail Criteria
Source Water Alkalinity
0--60
> 60 – 120
>120
Source Water TOC mg/L (Stream 1)
Removal %
Removal %
Removal %
>2.0 – 4.0
35
25
15
>4.0 – 8.0
45
35
25
>8.0
50
40
30
Table 2 Hardware Requirements
4195-1010-EPA
CRS
5 mg/L, 1 UV, 2 STREAM
4195-1020-EPA
CRS
10 mg/L, 1 UV, 2 STREAM
4195-1030-EPA
CRS
25 mg/L, 1 UV, 2 STREAM
4195-3010-EPA
SS
5 mg/L, 1 UV, 2 STREAM
4195-3020-EPA
SS
10 mg/L, 1 UV, 2 STREAM
4195-3030-EPA
SS
25 mg/L, 1 UV, 2 STREAM
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...