Page 80
Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
5.2.6.4 EPA Relay Mapping
The ability to map the pass/fail criteria, as defined above, is added
to each relay. This relay will hold the pass/fail criteria even when the
analyzer is offline.
5.2.6.5 Map Input
The STR 1 and STR 2 options apply
only to dual-stream analyzers.
An input SW1 is available to provide remote control. This volt-free relay
normally open contact can be used to trigger the following functions:
A three second contact closure will toggle the analyzer from one condition to
the other. The toggle back can only be done three seconds later. To enable
this function, enable the SW1 in the Service Menu.
5.2.7 Serial Menu
This option requires an optional communication module be installed in the
analyzer. See
1.
Enter the Serial menu from the Setup Menu.
2.
Press the UP button until the Serial screen is displayed.
3.
In the Serial screen, press ENTER. The Protocol screen is displayed.
5.2.7.1 Protocol Screen
There are 3 options under protocol:
•
CSV—outputs data logging and comma separated values.
•
ModBus—This is the Industrial Standard Protocol. More information
about the Modbus Protocol can be found on the Schneider
Automation website at www.modicon.com. The Modicon Modbus
S 1
* E P A P / F *
M a p i n p u t
S t o p / S t a r t
•
Start/Stop
•
Auto-clean/Go online
•
STR 2 Latched
•
Auto-validation/Go online
•
STR 1 Latched
•
Toggle Stream
•
Auto-calibration/Go online
S e r i a l
P r o t o c o l
C S V
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...