Section 5
Page 71
UVTOCSetup and System Configuration.fm
Setup Menu
Daily Mode
Select this mode only if the Val interval is set to zero hours. The Daily
operation sub-screens allow the selection by days. If the sub-screen entry is
set to Yes for Sunday and Friday the operation will occur automatically at the
operation time. If the sub-screen entry is set to No for Monday and Tuesday
the operation will not occur automatically those days.
The screens below set the analyzer to perform an automatic calibration cycle
23:00 only on days when the Daily calibrate daily entries are set to Yes,
(at least Sunday and Friday but not on Monday or Tuesday).
Disable operation
If the operation interval is set to zero, and all the daily entries Sunday through
Saturday are set to No, there will be no occurrences of that operation.
5.2.3.3 Blowback Off
The optional Blowback valve forces compressed air back through the inlet
filter to clear it of accumulated debris. The use of this option may not be
applicable in all applications.
1.
To enter the Blowback off screen, press the ENTER button after setting
the Clean interval.
2.
Set the minutes. To change the displayed setting, press ENTER.
3.
Asterisks will flash on either side of the setting. Use the UP or DOWN
buttons to increase or decrease the setting.
4.
When the time is displayed, press ENTER to place the value in memory.
The available intervals are 0 to 999 minutes. Setting 0 stops the
blowback function.
C a l t i m e
2 3: 00
C a l M o n d a y
No
C a l i n t e r v a l
o h r
C a l T u e s d a y
No
D a i l y c a l i b r a t e
C a l F r i d a y
Yes
C a l S u n d a y
Yes
B l o w b a c k o f f
3 0 m i n
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...