Page 140
Service Menus
UVTOCTroubleshooting and Service.fm
Section 8
In these screens HI/LO only allows
a change to be seen for the logical
level of the input. It does not refer to
the High or Low nature of the
parameter that is being monitored.
6.
Scroll through the following screens by pressing the UP and
DOWN button.
Disabling the carrier loss input will
enable pumps to run when the
carrier is not present. This can result
in serious damage to the carrier gas
system due to liquids back flowing
through the carrier gas components.
7.
To enable or disable a function, press ENTER.
8.
Asterisks will flash on either side of Enable or Disable. Press the UP and
DOWN buttons to toggle between Enable and Disable. Press ENTER to
save and store your selection.
9.
Press the UP and DOWN buttons to proceed to the next screen.
10. To exit the Input status function, press the CTRL then ENTER buttons.
You will be returned to the Input status menu.
8.6.3.1 Sparger and UV Flow Detection
The TOC analyzer is exposed to widely varying sample compositions.
Bubbles optically trigger the internal flow sensors when either a sparger or UV
flow failure occurs. In very difficult sample compositions, fluctuating optical
properties may cause a trigger of the bubble detectors even though no real
bubble occurs. In prior revisions of the firmware it has not been possible to
vary the sensitivity of these bubble detectors which has caused the analyzer
to either unnecessarily go offline to try and recover the liquid flow, or in the
case of dual stream analyzers, change to the alternate stream. For this
reason the ability to change the sensitivity of both the sparger flow and the UV
flow failure events has been added to the pre-existing Service/Input
Status/Sparger Flow and UV flow menus.
S p a r g e r f l o w
L O
15
U V f l o w
L O
15
R e a g e n t s
H I
E n a b l e
C a r r i e r l o s s
H I
E n a b l e
D r a i n l e v e l
H I
E n a b l e
S W 1
H I
E n a b l e
C e l l t e m p
5 0 . 0 C
E n a b l e
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...