Page 66
Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
5.2.1.4 Range Select Screen
The analyzer can operate so that a mapped analog stays in one of four
ranges or it can autorange. The ranges represent the behavior of the current
(4–20 mA) outputs, not the capability of the analyzer.
For example, if you have a 0 to 1000 mg/L analyzer:
You may choose to lock the analyzer into one of these fixed ranges or you
may use the full range (autoranging).
This will only be applicable only if an analog output is mapped to mg/L.
Autoranging:
If autorange is selected, the analyzer will automatically vary the 4–20 mA
mg/L signal as described in
. Enter the Autorange
screen from the Range change screen by pressing the UP button.
If you choose to autorange, map one of the analog outputs to one of the three
TOC output options and the other to Range ID. The analog output mapped to
Range ID will identify the present range of the analog output being used to
represent the mg/L reading.
If you have a 0 to 2000 mg/L analyzer, these range IDs are:
You must observe both analog outputs to determine the measurement.
For example, if output 1 is mapped to mg/L and the reading is 12 mA, and
output 2 is mapped to Range ID and the reading is 8 mA, the measurement
represents 250 mg/L.
Range 1
0 to 100 mg/L
Range 2
0 to 250 mg/L
Range 3
0 to 500 mg/L
Range 4
0 to 1000 mg/L
R a n g e s e l e c t
A u t o r a n g e
mA 1 range
mA 2 status
Range 1
0 to 200 mg/L
4 mA
Range 2
0 to 500 mg/L
8 mA
Range 3
0 to 1000 mg/L
12 mA
Range 4
0 to 2000 mg/L
16 mA
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...