Section 8
Page 135
UVTOCTroubleshooting and Service.fm
Fault Code Diagnostics
8.4 Fault Code Diagnostics
M12-M16
Not allocated
1
If this event persists, it may be disabled in the Service/Input Status Screen. See
Code
Fault Message
Diagnosis
Corrective Action
F011
No sparger flow
Loss of sample or acid
In cases where the sample composition is not
optically homogeneous, false triggering cannot
be avoided and the flow sensing should be
disabled in the Service\Input status screens.
In later firmware versions, this may require
setting the flow failure thresholds to 0.
Ensure no leaks in acid line due to either tubing
or fittings.
Verify sample is flowing to the analyzer.
F02*
No UV lamp flow
In cases where the sample composition is not
optically homogeneous, false triggering cannot
be avoided and the flow sensing should be
disabled in the Service\Input status screens.
In later firmware versions, this may require
setting the flow failure thresholds to 0.
Loss of persulfate.
Check persulfate supply.
Loss of resample.
Ensure no leaks in resample.
F03*
No carrier gas
The pressure switch has detected low carrier
pressure.
Verify carrier gas presence.
Replace pressure switch.
F04*
Code
Fault message
Diagnosis
Corrective action
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...