Section 3
Page 51
UVTOCReagents and Standards.fm
Preparing Phosphoric Acid Solution
Caution: Always pour the
concentrated acid into the water;
never pour water into the acid.
Take the same precautions you
would normally take when
handling any corrosive material.
for
Spanish, French, German, or
Italian translations.
The ideal sample pH after addition of acid is 3.0 or less, for efficient removal
of inorganic carbon. In some cases, stronger solutions and/or higher pumping
rates may be required to adequately maintain the pH of the sample in the TIC
scrubber. Determine the necessary acid reagent concentration by adding
phosphoric acid to a sample until the pH is reduced to 3.0 or less. To verify the
pH is 3.0 or less, collect a sample from the drain line.
details the concentration of acid required to remove 10 mg/L and 100
mg/L C as TIC for each analyzer. Select the appropriate molarity to remove
the anticipated TIC in your sample.
Table 7 Reagent Tables
UV 4195- / 6195-
1010/
3010
1020/
3020
1030/
3030
1040/
3040
1050/
1050
1060/
3060
1070/
3070
2000/
4000
TOC Range mg/L
5
10
25
50
100
200
500
100
Sodium Persulfate
Molarity
0.2
0.4
0.6
0.6
0.4
0.5
0.6
0.7
g per 20 L
950
1900
2860
2860
1900
2380
2860
3330
Usage days per 20 L
90
90
90
90
90
90
90
30
Phosphoric Acid
TIC mg/L C (Na
2
CO
3
)
10
100
10
100
10
100
10
100
10
100
10
100
10
100
10
100
Molarity
0.1
0.6
0.1
0.6
0.1
0.6
0.05
0.3
0.02
0.1
0.02
0.1
0.02
0.1
0.1
0.6
mL (85% H
3
PO
4
) per
20 L
137
819
137
819
137
819
68
410
27
137
27
137
27
137
137
819
Usage days per 20 L
30
30
30
30
30
30
30
30
UV 4195-
2010/
4010
2020/
4020
2030/
4030
2040/
4040
2050/
4050
2060/
4060
2070/
4070
2080/
4080
TOC Range mg/L
200
500
1000
1000
2000
5000
10000
20000
Sodium Persulfate
Molarity
1.4
1.6
0.9
0.9
1.5
1.2
0.9
1.2
g per 20 L
6660
7620
4280
4280
7140
5710
4280
5710
Usage days per 20 L
30
30
30
30
30
30
30
30
Phosphoric Acid
TIC mg/L C (Na
2
CO
3
)
10
100
10
100
10
100
10
100
10
100
10
100
10
100
10
100
Molarity
0.1
0.6
0.04
0.3
0.04
0.3
0.04
0.1
0.02
0.1
0.02
0.1
0.02
0.1
0.02
0.1
mL (85% H
3
PO
4
) per
20 L
137
819
55
410
55
410
55
137
27
137
27
137
27
137
27
137
Usage days per 20 L
30
30
30
30
30
30
30
30
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...