5.2.8.2 Span Tolerance Screen
Span tolerance is acceptable CO
2
ppm variation (±) allowable when an
autocalibration span operation is being performed.
Press the Up button. The Maint% Pri screen is displayed.
5.2.8.3 Maint % Pri Screen
This percentage value is used as a maintenance event threshold indicating
a change in analyzer performance since a primary calibration. For example,
to indicate that the peristaltic tubing on the analyzer pumps may need to be
replaced. If the current calibration differs from the primary calibration by
more than this percentage, a warning maintenance event will be identified.
Refer to
for a detailed discussion of maintenance
event codes.
5.2.8.4 Fault % Pri Screen
This percentage value is used as a fault event threshold, indicating change in
analyzer performance to indicate that the current calibration is too far from the
primary calibration to be acceptable, and that the TOC reading is not
necessarily correct. If the current calibration differs from the primary
calibration by more than this percentage, a fault maintenance event will be
annunciated. Refer to
fault codes.
5.2.8.5 Averaging Screen
This screen allows entry to a menu allowing the averaging to be changed.
The greater number of readings that are averaged, the smoother the plot.
Enter the passcode before accessing these settings.
The IR gas averaging screen is displayed.
S p a n t o l e r a n c e
1 0 p p m
M a i n t
% P r i
1 0 %
F a u l t
% P r i
15%
A v e r a g i n g
I R g a s
* A d a p t i v e *
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...