
Page 86
Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
Press ENTER when the language of choice is displayed; your selection will be
placed in memory.
5.2.10 Analyzer Screen
The Analyzer screen permits you to select the model analyzer you have.
The analyzer number corresponds with the instrument's analysis range.
Before changing an analyzer model, record the Zero and Span counts for
Analog Channels CH1 and CH2. Access these menus in the Service/Output
adjust submenus, see
Selecting a new configuration
will force a reset of the
analyzer software, and will overwrite
the calibrations with default
calibrations and setup parameters.
1.
To enter the Analyzer screen, press the UP button while in the Language
screen. The Analyzer screen is displayed.
2.
The analyzer will prompt for the passcode twice before changing the
analyzer model in this screen. See
for passcode
information.
3.
To change the analyzer model number, press ENTER. Asterisks flash on
either side of the model designation. Press the UP or DOWN buttons to
scroll through the various model numbers.
4.
When your analyzer model is displayed, press ENTER. This places your
selection in nonvolatile storage, and will reboot and load the default
settings for the selected analyzer.
5.
Reenter the CH1 and CH2 Zero and Span settings in Service/Output
adjust.
6.
Press CTRL then ENTER to return to the Setup Menu.
In most situations, when a different analyzer is selected it will be necessary to
reconfigure hardware. See
.
5.2.11 EPA Main Operating Screen
A dual stream analyzer hardware
configuration is required for EPA
operation.
In EPA mode the screen cycles between EPA Setup, Retention Time, and
Alkalinity. The screen relating to the stream currently selected is held for
6 seconds, while the other two screens are held for 3 seconds. The screen
relating to the unselected stream displays the time the unselected stream
was last measured.
See
for additional manual control screens.
5.2.11.1 EPA Setup Screen
The selection of the EPA application will enable the EPA setup menu option,
positioned below the Manual control screen. Two user entry screens are
available to input the Retention Time in hours and minutes, and Alkalinity
concentration in mg/L.
A n a l y z e r 4 1 9 5 -
1 0 0 5 / 3 0 0 5
E P A s e t u p
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...