Section 2
Page 45
UVTOCUnpacking Installation Setup.fm
Serial Data Connections
Figure 24 RS485 Connection to Remote Device
Important Note: Install a jumper from JP1-2 to JP1-3 (lower two pins) on the I/O pcb
(130143) for RS422 and RS485 communications (
2.5.2
Optional Blowback Filter Valve
The optional blowback filter valve is controlled by an internally-fused, switched
relay output supplying 24 V dc (maximum 500 mA) to a valve on an extended
blowback filter system. This output may be configured for separate ON and
OFF periods. The ON period is configurable between 1 and 100 seconds, and
the OFF period is configurable between 1 and 999 minutes.
This output is designed to supply power to only one blowback filter valve.
Interposing a relay and power supply will be required if more than 0.5 A is
required for valve(s) operation.
Installation information is supplied with the blowback filter valve.
Figure 25 Locating the Blowback Filter Connector
1.
8001 I/O Printer Circuit Board Assembly
2.
Blowback Filter Connector
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...