Page 134
Maintenance Code Diagnostics
UVTOCTroubleshooting and Service.fm
Section 8
M06
Validation %
The analyzer failed to verify that the sample connected to
the standard port was within the % of the user specified
validation standard. A calibration may follow this failure.
See
1.
Check the standard is the correct mg/L.
2.
Recalibrate the analyzer.
3.
Increase the validation %.
4.
Verify the user entered standard concentration.
M07
Calibration failure
The current calibration deviates from the primary
calibration by greater than the Fault % Primary entry in
the Setup/ Tolerance/Fault % Primary Screen. See
1.
Replace peristaltic pump tubing and perform a
primary calibration.
2.
Edit the entry in the Setup/Tolerance/Fault % Pri
Screen.
M08
Reactor Flow
The pressure in the UV reactor has exceeded the user
specified value in Maintenance Pressure Threshold
Setup/Tolerances.
The reactor flow is being restricted. Locate and remove
the cause of the restriction. Possible causes are:
1.
UV manifold crystal buildup in the UV block. Clean
the UV manifold.
2.
Blocked UV reactor. Disassemble the reactor and
flush with water. If the blockage is not removed with
flushing, contact technical support.
M091
Low reagents
One of the inputs to the analyzer has detected a
low level.
1.
Refill the bottles.
2.
Verify jumper settings
M10*
Low cell temperature
1.
Cell heater plug not connected to 8001 I/O PCB
assembly.
2.
Cell heater faulty.
1.
Connect IR cell heater. Determine of IR cell is warm
to the touch.
2.
If LED is cycling on and off, replace the cell heater.
3.
If erratic cell temperature is experienced, stop the
analyzer. If the cell temperature subsequently
becomes stable, then it is likely that either a UV
power supply or a UV lamp is defective, causing high
EMI to affect the temperature reading.
M11
Spill detection
The TOC reading has achieved a % level of range in
excess of the Spill Threshold %, and the analyzer has
gone into a clean cycle for the spill recovery time.
See
1.
To disable this feature raise the spill threshold
percent.
Code
Fault message
Diagnosis
Corrective action
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...