Page 78
Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
5.2.6.3 Map Relays
Mapping events to relays allows the user to customize the system. In addition,
the relay state, either energized or deenergized, may be selected. If relay S1
is mapped to Alarm 1 and selected as energized, the relay coil will be
energized when Alarm 1 occurs.
Multiple events cannot be mapped
to one relay, however an event can
be mapped to multiple relays.
When the UP button is pressed after mapping the channels, the Map relays
screen is displayed.
Press ENTER. The Screen to map S1 relay is displayed.
Relay S1 may be mapped to:
Alarm 1—a rising level nonlatched alarm.
Alarm 2—a rising level nonlatched alarm.
Online—the Online status of an analyzer indicates that the analyzer analog
outputs (4-20 mA) are representative of the current TOC reading. When
mapped to an output relay, the Online status allows DCS systems to
determine that the 4-20 mA is representative of the current TOC reading. This
is necessary because if the analyzer is not currently measuring the sample
stream the 4-20 mA output may be latched to the last valid TOC reading when
it was measuring the sample. The Online relay will not be in its TRUE status
during this time. Furthermore, when the analog output has been mapped to
Live mg/L, the output will reflect whatever is being analyzed.
Offline—activated when the analyzer is not online, such as when it is
warming up, purging, cleaning, calibrating, validating, or providing a single
sample reading.
Maintenance—activated when any of the maintenance events occurs, such
as low reagents, and zero stability. These are detailed in
Fault—activated when any of the fault events occur, such as no sparger flow,
no UV lamp flow, no carrier gas. Faults are detailed in
C H 2
* L a t c h e d m g / l 2 *
R e l a y S 1
A l a r m 1
M a p R e l a y s
R e l a y S 1
A l a r m 1
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...