Page 144
Service Menus
UVTOCTroubleshooting and Service.fm
Section 8
3.
When finished, press ENTER to place the setting in memory.
4.
Press the UP button. The Span screen with the Span digital count is
displayed.
Channel 1 and 2 Span
The value of 3890 is the current span digital count for the selected output
channel (CH1 or CH2). Verify that 20 mA are displayed on your multimeter.
1.
To adjust the displayed value and the current (mA), press ENTER.
2.
To accept the reading, press ENTER.
3.
Use the UP and DOWN buttons to ramp the value up or down. The count
range for Span is 1025 to 4095 counts. The displayed count must equal
20 mA on your multimeter or the appropriate TOC value on a DCS.
Channel 1 and 2 Cal Test
To enter the Channel test screen, press the UP button while in the Channel
select screen.
The purpose of Cal test is to verify that the there is linearity over the 4 -20 mA
range. The 50% value shown above should equal 12 mA on your multimeter.
A value of 0% equals 4 mA and a value of 100% equals 20 mA.
8.6.6 Setting the Elevation
This menu allows the IR to be calibrated at a different elevation, and thus
different atmospheric pressure, to the final operating elevation of the analyzer.
It will normally be set to 1m at elevations at or close to sea level. If the
analyzer is being operated at substantial elevations using a factory calibrated
IR it will be necessary to set this menu to this local elevation. If the analyzer is
gas calibrated on site in accordance with the instructions in this manual,
entering the local elevation is not necessary, and it should be set to 1 m.
8.6.7 UV Block Selection
In order to get the correct default bubble threshold constants for each
analyzer range, a new menu, Service/UV Block, has been added. This menu
allows selection of the block by part number. All of the original acrylic blocks
can be treated as the part number 011664. The new block part number is
7121900. The default selection in this release of firmware is the old
block, 011664.
To be effective, the UV Block menu must be selected after the analyzer is
selected. The default values can be overridden by the Sparger Flow and UV
Flow menus as detailed in
.
C H 1 s p a n
3890
C H 1 t e s t
50%
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...