Introduction
Section 1
Introduction
1.1 System Description
See
for a
list of abbreviations, acronyms, and
definitions used in this manual.
This manual applies only to Cat. No. H-4195; –X010; –X020; –X030; –X040;
–X050; –X060; –X070; –X080 and H-6195– analyzers and describes the
installation and operation of the analyzer, shown in
. The
analyzer measures organic carbon in liquid samples ranging from 0.015 mg/L
to 20,000 mg/L. In the analysis process, carbon is converted to
carbon dioxide, which is then measured with an IR detector. The analyzer
uses the EPA approved UV-persulfate oxidation method, which produces the
most reliable TOC measurement. The analyzer meets customer specifications
worldwide, and complies with ISO, EPA, EN, DIN, CE, UL, CSA, FM,
and NAMUR regulations.
Software navigation is simplified by the use of dialog prompts for single
sample analysis, analyzer validation, and analyzer calibration.
The analyzer is capable of local 4–20 mA, optional RS232 CSV data logging,
and ModBUS communication.
The analyzer is designed to facilitate customer maintenance. All key
components are located for easy access. All connection ports are located on
the outside of the analyzer. A dual-stream option is available.
Sample dilution can be added to the standard instrument. The optional
dilution configuration is shown in illustrations throughout this manual.
1.2 Applications
The analyzer can measure either Total Carbon (TC) or Total Organic Carbon
(TOC) in water continuously or from grab samples. It is recommended for use
in the following applications:
•
Boiler feed water
•
Bulk chemical
•
Pharmaceutical process water
•
Condensate and cooling water
•
Unit out fall
•
Industrial wastewater
•
Industrial effluents
1.3 Major Components
There are two enclosures—one for liquids and one for electronics.
The enclosures are available in epoxy powder coated steel or stainless steel.
The top electronics enclosure is separate and sealed from the bottom liquids
enclosure. Refer to
.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...