Section 1
Page 17
UVTOCIntroduction.fm
Major Components
1.3.1 Liquids Enclosure
This enclosure houses the inlet, sparger, and reactor manifold. In addition, it
contains the gas-liquid separator (GLS), the pumps, the UV assembly, and the
liquid drains.
•
Fluid flow is simplified by using three manifold assemblies:
•
The inlet manifold contains all the connections for reagents, stream,
and calibrations standards, and associated valves
(
).
•
The sparger manifold is integrated with the flow control and pressure
gauge. It provides the mixing of the acid and sample as well as the
liquid flow detection. See
•
The reactor manifold provides the mixing of sample and persulfate
(
).
•
Two pump motors are used for the entire analyzer range. Different pump
heads are used to obtain different ranges (
).
Some optically inhomogeneous
sample compositions will provide
false triggering, requiring that the
flow failure be disabled in the
Service\Input Status screen.
•
The UV reactor assembly is capable of mounting up to four standard UV
lamps and their encapsulated electronic switching power supplies
(
).
•
Flow detection is available as sample and reactor flow using bubble
detection, which is fed into the electronic assembly for appropriate action.
1.3.2 Electronics Enclosure
This enclosure contains the IR bench, the 8000 Display/Controller PCB
assembly, the 8001 I/O Termination PCB assembly, and the configurable
115/230 V 50/60 Hz analyzer power supply. The front panel of this enclosure
contains the window for the 2 line x 16 character alphanumeric display, three
analyzer status LEDs, and the four-button tactile membrane keypad (
and
Additional information on the front panel is contained in
.
•
The 8000 Display/Controller PCB assembly is factory configured to
an IR range of either 1,000 or 10,000 ppm CO
2
and a liquid range
between 0–5 mg/L and 0–20,000 mg/L TOC. Connection to the IR
block uses a 25-way ribbon D connector. Connection to the 8001 I/O
is via a ribbon cable.
•
The 8001 I/O Termination PCB assembly provides the connection
point for five function-mapped relays and/or two parameter-mapped
4-20 mA analog outputs.
•
The power supply, from an input of 115 V ac or 230 V ac, provides
115 V ac for the pumps and 24 V dc for the rest of the electronics. The
power supply is shown in
.
All user connections inside the electronics enclosure are screw
down terminals.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...