Page 64
Setup Menu
UVTOCSetup and System Configuration.fm
Section 5
5.2.1.1 Alarm Levels 1 and 2 Screens
The first level alarm is a rising level unlatched alarm. At this level, alarm
output 1 goes into an active state when the mg/L concentration exceeds the
setpoint. When the alarm is activated, the front Panel LED (A1) will illuminate.
Alarm 1 and Alarm 2 both apply to
the stream currently being
measured.
If the alarm has been mapped to a relay, the relay will reflect its active state.
1.
To change the setting, press ENTER.
2.
You will be asked to enter the passcode (1953). After entering the
passcode, press ENTER.
3.
Asterisks will flash on either side of the displayed alarm value. When the
asterisks are flashing, you can modify the alarm level by pressing the
ENTER or DOWN buttons.
4.
Enter the desired value.
5.
When the desired value is displayed, press ENTER to accept the new
alarm level.
The second level alarm is also a rising level unlatched alarm.
Like Alarm 1 level, Alarm level 2 goes into an active state when the mg/L
concentration exceeds the setpoint. To access the Alarm 2 level screen, press
the UP button. Set the value as above for Alarm 1.
If no alarms are required, set the values high to prevent any actions.
5.2.1.2 Inhibit Level
The inhibit level is a TOC percentage of the analyzer’s full scale range.
If the inhibit level is exceeded at the time of the scheduled calibration or
clean cycle, the cleaning, calibration, and validation cycle is prevented from
occurring. Use this feature to continue monitoring a high carbon concentration
even if an automatic clean or calibration is scheduled. Disable this feature by
entering 0%.
1.
Enter the Inhibit level screen by pressing the UP button while in the Alarm
screens. The display will be similar to the following.
2.
To change the displayed setting, press ENTER.
3.
Asterisks will flash on either side of the displayed value. Use the UP and
DOWN buttons to select the new value.
4.
Press ENTER when finished to store the new Inhibit level in memory.
A l a r m l e v e l 1
3 . 0 0 m g / l
I n h i b i t l e v e l
80%
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...