Page 46
Carrier Gas and Fluid Connections
UVTOCUnpacking Installation Setup.fm
Section 2
2.5.3
Optional Level Detector Configuration
Optional level detectors may be installed to warn of low reagent levels or a
high wastewater level in the drain pan.
shows connections and
jumper configurations that are made on the I/O PCB. A signal from a sensor
indicates a low reagent level.
If a third reagent level detector is used, remove the jumper from LK12 and
install the sensor on terminals 11 and 12.
Figure 26 Optional Level Detector Connections and Setup
2.6 Carrier Gas and Fluid Connections
2.7 Connection Information
Accommodate s
amples between
70–100 °C (160–212 °F) using
an external cooler consisting of
a 3000 mm (120-in.) long, 6 mm
(¼-in.) O.D. stainless steel
sample inlet tube at a flow rate
of 25–60 mL/minute.
Carrier gas and fluid connections are made on the lower left-hand side of the
analyzer. Refer to
. After connections are made, ensure
that they are leak free.
On non-dilution analyzers a connection to the dilution port is not necessary.
Plug the dilution port.
If the analyzer is not equipped with a fast-loop sweep, the sample port may be
connected to an ambient pressure source.
If the analyzer is equipped with a fast-loop sweep, the sample entry port can
be connected to a pressurized source of up to 6 bar (87 psi), with a control
valve (needle valve) placed on the inlet to throttle down the flow to between
25 and 200 mL/min.
1.
Reagent Level Detector 1 connected to REA 1
3.
Internal Level Detector
2.
Reagent Level Detector 2 connected to REA 2
4.
LK12
Important Note: A jumper is installed on LK12 when the Optional Level Detectors are connected. If a Reagent Level Detector
is removed, a jumper must be placed on the corresponding link, (LK10 = REA 1, LK11 = REA 2).
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...