Page 116
IR Cell Cleaning
UVTOCMaintenance.fm
Section 7
Figure 38 Installing a Pump
7.6
IR Cell Cleaning
If relay contacts are connected to ac line voltage, remotely disconnect
the power before opening the analyzer’s upper enclosure door.
before performing this procedure.
The IR cell assembly cleaning interval is application dependent. Minimally,
it should be cleaned once a year. If analysis results become unstable, it may
be an indication that the IR cell needs cleaning. For the purposes of clarity,
the accompanying illustrations show the IR cell removed from the analyzer.
Required cleaning materials:
•
Isopropyl alcohol (clean water is also effective)
•
Lint-free cleaning tissues
•
Long cotton swabs
•
Cutting pliers (to cut the tie wrap holding the thumb latch to allow the
cell assembly to be removed from the IR bench).
1.
Refer to
and remove the inlet and outlet tubing
from the IR cell.
2.
Support the mirror assembly (to prevent the cell assembly from falling out
if the thumb latch unexpectedly releases) and cut the tie wrap.
3.
Release the thumb latch by pulling it away from the cell assembly.
4.
Clean the interior wall of the cell assembly with lint-free cleaning tissues
and isopropyl alcohol, as shown in
See Safety Note 40 on page 118
5.
Use cotton swabs and isopropyl alcohol to clean the sapphire window that
protects the parabolic mirror, as shown in
.
6.
Inspect the assembly to make certain that all debris and contamination
are fully removed and that the sapphire is free of scratches and cracks.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...