Page 130
Troubleshooting
UVTOCTroubleshooting and Service.fm
Section 8
used for this. Reconnect reagents and start the analyzer. The analyzer
should now respond to the standard.
5.
The delivery rates (mL/min) of all pumps can be verified against the
hardware configuration table using water instead of acid and persulfate.
This should not be done with acid and persulfate in the analyzer. Use a
small measuring cylinder to measure how much water is taken from the
cylinders. This is more accurate than measuring the delivery from the
pump, because undetected downstream blockages (not in the flow path)
could affect the results.
8.2.2 Instrument is Unstable or Drifts
1.
Check for proper installation of electrical input wiring.
2.
Check nitrogen flow rate for stability.
3.
Check gas calibration (see
).
4.
Pinched or incorrectly installed pump tubing.
8.2.3 Periodic Spikes in the Readout
Verify that large particles are not being introduced into the reactor. Large
particles in the sample will cause periodic spikes in the readout. Homogenize
and/or filter the sample adequately. At elevated temperatures, persulfate
bubbles can also cause dips followed by spikes in the output, do not exceed
the molarity specified in
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...