Section 5
Page 85
UVTOCSetup and System Configuration.fm
Setup Menu
The IR gas channel has five filter settings to produce different
moving averages:
For this analyzer, Adaptive is the preferred setting.
1.
To change the filter setting, press ENTER.
2.
Asterisks will flash on either side of the displayed setting. When the
desired filter setting is displayed, press ENTER to store the selection in
memory.
5.2.8.6 Maintenance Pressure
A pressure sensor on the 130143
PCB measures the pressure at the
UV block. When the pressure rises
above the threshold pressure, the
analyzer will automatically stop the
pumps, preventing the carrier gas
system from being damaged by the
corrosive liquids in the analyzer.
This is the threshold above which the maintenance event for reactor pressure
will occur.
5.2.8.7 Fault Pressure
This is the threshold above which the fault event for reactor pressure
will occur.
5.2.9 Language Screen
The Language screen is entered by pressing the UP button while in the
Tolerances screen.
1.
To change the displayed language, press ENTER.
2.
Asterisks flash on either side of the displayed language.
3.
Press the UP or DOWN buttons to scroll through the language selections.
The choices are English, Francais, Deutsch, Espanol, and Italiano.
Setting
Sampling mode
Low
4 samples every 2 seconds
Medium
20 samples every 10 seconds
High
40 samples every 20 seconds
Very high
80 samples every 40 seconds
Adaptive
Adaptive filter
M a i n t P r e s s u r e
6 0 k A
F a u l t P r e s s u r e
1 0 0 k P a
L a n g u a g e
E n g l i s h
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...