Section 6
Page 95
UVTOCOperation.fm
Validation Menu
Two additional screens lead to Service and Setup menus. To access these
from the main menu, press CTRL and DOWN for the service screen and from
the Service screen, press the UP button for the Setup screen.
Definition of a screen: An item within a menu.
6.8.1 Passcode Protection
When entering into passcode
protected menu items, the operation
performance and/or reliability of the
analyzer might be altered or
damaged inadvertently. Only
experienced users should access
these menus.
Some menu functions are passcode protected. Entering a passcode is
required before editing can take place. However, all screens may be viewed.
The system passcode is 1953. Use the UP or DOWN buttons to scroll to
1953. Then press ENTER. Access to any of the functions is available until the
menu tree is exited.
The passcode is fixed and cannot be changed.
6.9 Validation Menu
Assess the validity of the current calibration by performing a validation of
a known TOC standard. This function may be initiated manually or
automatically scheduled.
After the TOC value of the known standard has been entered, the standard is
introduced via the Calibration port and then analyzed. After the reading has
stabilized, the result is displayed before proceeding online.
Auto validation uses the TOC standard concentration specified in the
Calibration/Autocal/Val Std. screen.
The result is displayed alternately as a TOC mg/L reading and a percentage
of the entered standard. The result is stored and can be viewed until
overwritten by the next validation. The result of the validation as a percentage
is compared to a validation tolerance. If the validation result deviated by more
than the validation tolerance, a validation event occurs. In addition, if a Cal
after Val is selected, a calibration will follow the failed validation.
View the result of the prior validation by pressing ENTER.
E n t e r p a s s c o d e
* 1 9 5 3 *
V a l i d a t i o n
2 / 1 4 / 2 0 0 0
05 :3 4
V a l
9 7 0 m g / l
2 / 1 5 / 2 0 0 0
12 :3 8
V a l 9 9. 7%
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...