Section 5
Page 69
UVTOCSetup and System Configuration.fm
Setup Menu
5.2.2.6 Clean Options
This submenu allows entry into two clean-related options — Clean with lamps
on and Clean before calibration.
Danger: Turn the lamps off
before cleaning. Cleaning with
some solutions with the lamps on
will produce toxic substances.
See Safety Note 10 on page 185
for Spanish, French, German, or
Italian translations.
To enter the two clean related options, press ENTER. Select Yes or No for
each option.
Configure the analyzer to perform a clean operation with the lamps on or off.
Lamps should be switched off if the cleaning solution (such as bleach) could
produce toxic vapors (such as chlorine gas).
If Yes is detected in the Clean before cal screen, a cleaning operation will
always occur before an autocalibration. Cleaning before calibration is
recommended.
5.2.2.7 Autocal Options
These screens determine if a zero only, a span only, or both a zero and span
calibration adjustment are performed during an auto calibration.
Edit the selection using ENTER and UP & DOWN buttons.
Zero and Span: Both a Zero and a Span are performed.
Zero only: Only a Zero is performed.
Span only: Only a Span is performed.
5.2.3 Timing Screen
The Timing screen is entered from the Cal to use screen by pressing the UP
and DOWN buttons.
C l e a n o p t i o n s
C l e a n w i t h
l a m p s o n ?
No
C l e a n b e f o r e c a l
Yes
A u t o c a l o p t i o n s
C a l i b r a t i o n
Z e r o & S p a n
T i m i n g
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...