Section 4
Page 59
UVTOCCommissioning.fm
Analyzer Adjustments
8.
Place the sample stream line back into DI water.
9.
Allow the reading to stabilize to lower than 30 ppm CO
2
.
10. Enter the Calibration/Primary Zero menu and follow the
prompted procedure.
11. Visually ensure that the CO
2
ppm is stable. Because the analyzer has
already analyzed DI water, press ENTER repeatedly to skip the elapsing
time and advance to the final screen.
12. Go back to the Main screen.
13. Connect the sample stream line to the sample source.
14. Set-up the local parameter settings such as Auto-cal, Auto-clean, Relays,
Analog outs, etc. as described in the instrument manual.
Figure 29 Bottom View of the Pump Bracket Assembly
1.
Flow Adjustment Knob
2.
Pressure Gauge
3.
Pressure Regulator Knob (Turn clockwise to decrease pressure. Turn counter-clockwise to increase pressure.)
Note: To access the pressure regulator, carefully open the pump module without pulling any tubing off of the barb fitting.
1
2
3
Front of
analyzer.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...