Section 7
Page 119
UVTOCMaintenance.fm
Pressure/Leak Test
7.7 Pressure/Leak Test
If relay contacts are connected to ac line voltage, remotely disconnect
the power before opening the analyzer’s upper enclosure door.
before performing this procedure.
The pressure/leak test reasonably verifies that the O-rings have seated and
that they are sealing the IR cell assembly. To perform this test:
1.
Remove the outlet tube from the IR cell assembly and plug the outlet tube.
Refer to
.
2.
Observe that the bubbles in the GLS U-tube consistently travel from front
to back. If they do, the cell O-rings are providing a sufficient seal.
Failure of this test may also indicate a cracked condenser or leaking
GLS fitting.
Figure 43 IR cell plugged during the leak test
7.8 IR Cell Replacement
Disconnect all power sources to the analyzer and to the relays before
replacing the IR Cell.
7.8.1 Thumb Latch Adjustment
After an IR cell is replaced, improper
adjustment of the latch can damage
the new cell assembly and make
mirror alignment impossible.
1.
Remove the IR cell by releasing the thumb latch.
2.
Loosen the retaining nut on the piston latch. Refer to
and
.
3.
Place the IR cell in the optical bench, align the pin(s) and move the latch
to engage the piston in the indentation on the end of the cell.
Plug this port as shown here.
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...