Section 5
Page 67
UVTOCSetup and System Configuration.fm
Setup Menu
Fixed range:
If you choose fixed range and either or both of the analog outputs are used to
output the mg/L, the analyzer will have a fixed 4–20 mA mg/L span.
1.
To make a range selection, press UP while in the Range change %
screen. The display changes to:
2.
Press ENTER. Asterisks will flash on either side of Auto range.
3.
To accept Auto range as your selection, press ENTER again.
4.
To select Fixed Ranges 1 through 4, press the UP button. Each time you
press the button, the next highest range is displayed.
5.
The asterisks continue to flash on either side of the available ranges.
To select a displayed range, press ENTER. The selection will be placed
in memory.
5.2.1.5 Selecting Output Ranges
Allowing the analog output scaling to change effectively allows
intermediate ranges to be presented to a DCS or monitoring system.
Selecting output ranges refers only to the mapped analog outputs when they
are mapped to TOC.
It is not possible to adjust an analog
output range to a value greater than
the real range of the analyzer.
1.
After you have selected a fixed range, press the UP button. The Output
ranges screen is displayed.
2.
Press ENTER. The Output range value screen is displayed.
3.
Press the UP and DOWN buttons to scroll through the four ranges.
To change a range value, press ENTER.
4.
Asterisks will flash on both sides of the displayed value. Use the UP and
DOWN buttons to change the value. When finished, press ENTER. The
new value is placed in memory.
5.
Press CTRL then ENTER to exit the Output range screen.
O u t p u t r a n g e s
O u t p u t r a n g e s
O u t p u t r a n g e 1
1 0 0 . 0 m g / l
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...