Analog Outputs
2.4
Analog Outputs
Only Jumpers LK4, LK5, LK7, and LK8 relate to the analog outputs. The
default configuration on the I/O PCB assembly is 4–20 mA, no links (jumpers)
installed. These connections are labeled CH1 Analog and CH2 Analog.
Although isolated from the rest of
the electronics, the two 4–20 mA
circuits are not isolated from
each other.
Use shielded twisted-pair signal cable with the shield grounded at the receiver
end. Connect the positive signal lines to the I+ terminals and the negative
signal lines to the I– terminals. The maximum load for each analog output is
600 ohms. For on-site calibration of the 4–20 mA, see section
. See
Technical Specifications on page 13
for resistive load requirements.
The 4–20 mA analog outputs are
self-powered. The resistive load
should have no other
voltage connections.
Optionally, using links LK4, LK5, LK7, and LK8 (as described in
the analyzer to be set for 0–10 V output or 4–20 mA output with
I– connected to ground (PE).
These settings are also detailed on the I/O PCB.
2.5
Serial Data Connections
Connection cables are user supplied and must be shielded to minimize
electromagnetic effects on analyzer performance. Refer to
Figure 19, Proper Wire Preparation and Insertion, on page 40
.
Using this option requires an RS232
or RS485 Kit. See
for catalog
numbers.
The Serial Data Communication connection allows an optional kit to be
installed in the analyzer providing optically isolated user serial
communications capability to the analyzer. Both CSV data logging and
ModBUS communications are supported, allowing connectivity to simple and
complex communications equipment.
To ensure that ground currents in inadequate ground systems do not result in
potential shock hazards, connect the shield at only the analyzer end. When
wiring the other end, cut the shield wire off and remove it. Do not connect the
shield wire at both ends.
Table 4 Link Setting For Analog Output
Link
4–20 mA
10 Volt
I– to PE (ground)
CH1
CH2
CH1
CH2
CH1
CH2
LK4
Yes
LK8
Yes
LK5
Yes
LK7
Yes
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...