Page 108
Manual Control for Selecting Stream 1 or 2
UVTOCOperation.fm
Section 6
6.12.5 Inhibit Next Clean
Inhibiting any scheduled operation
will only prevent the occurrence of
the next scheduled operation, not
subsequent operations.
Press ENTER to Inhibit next clean. An Inhibited confirmation screen is
briefly displayed.
6.12.6 Inhibit Next Calibration
Press ENTER to Inhibit next calibration. An Inhibited confirmation screen is
briefly displayed.
6.12.7 Inhibit Next Validation
Press ENTER to Inhibit next validation. An Inhibited confirmation screen is
briefly displayed.
6.12.8 Active Calibration Set Selection
It is possible to select one of six calibrations as the current basis for
calculation of TOC: Manual, Primary, Auto-cal, Segmented, TOC added, and
Fitted. To change the displayed selection, press ENTER. See
.
6.12.9 Reboot System?
Press ENTER to reboot the system. In effect, the analyzer is reset without the
need to turn it off and back on.
When the system is reset, the previous settings are retained and the Main
Operating screen is displayed.
If you do not want to reset the system, exit the screen by simultaneously
pressing the ENTER + CTRL buttons.
6.13 Manual Control for Selecting Stream 1 or 2
Once dual stream has been selected in the setup menu, an extra screen
is inserted into the Manual control menu. This allows manual switching
between the two streams.
I n h i b i t n e x t
c l e a n ?
I n h i b i t n e x t
v a l i d a t i o n ?
A c t i v e c a l
A u t o c a l
R e b o o t s y s t e m ?
C h a n g e t o
S t r e a m 2
C h a n g e t o
S t r e a m 1
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...