Section 8
Page 143
UVTOCTroubleshooting and Service.fm
Service Menus
8.6.5 Output Adjust
The procedure to change any of the
displayed values is identical for
Analog channels 1 and 2.
The Output adjust function allows calibration and checking of the two analog
channels. It also allows the output of a % scale for the selected output
channel by first selecting the channel and then selecting the zero, span, or
check option.
The analog outputs are line
powered and should not be
connected to a load with voltage
that is independently applied.
The output adjust menu is particularly useful to trim the analyzer’s output to
exactly match the input of a DCS system—allowing accurate repeating of
signalled values.
In order to accurately calibrate the analog outputs, the two analog circuits
must be connected to the circuit loop which operates analyzer for at least five
minutes before any adjustments are performed. This means that the loop
must not be broken (open-circuited) during the adjustment of zero and span
with the DCS loop load resistor is in place. If the jumper configuration is
changed recalibration may be required. For wiring details and jumper
configuration, see
. For electrical parameters, see
Technical Specifications on page 13
.
1.
To display the Output adjust screen, press the UP button while in the
Service menu.
2.
Press ENTER. The CH1 (Analog channel 1) screen is displayed. To
display CH2 outputs, press the UP button. Press ENTER to select the
displayed channel.
The three options for each channel are Zero, Cal, and Cal check. (See the
complete descriptions of each option below.) Scroll through the options using
the UP and DOWN buttons. Press ENTER to select the option.
Channel 1 and 2 Zero
The value of 608 is the current zero digital count for the selected output
channel (CH1 in the screen above). Verify that 4 mA are measured on your
multimeter or that 0 is repeated on a DCS.
1.
To adjust the reading, press ENTER.
2.
Asterisks flash on either side of the displayed value. Use the UP and
DOWN buttons to ramp the value and the current (mA) up or down. The
count range for Zero is 0 to 1024 counts. The displayed count must equal
4 mA on your multimeter.
O u t p u t a d j u s t
C H 1
C H z e r o
608
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...