Replacing the Fuses
7.13 Replacing the Fuses
D
Danger: Remove power to the
analyzer if any electrical work is
to be performed.
for Spanish,
French, German, or Italian
translations.
The analyzer contains a total of six fuses. See
for their location. and
the text below for their specifications.
Remove power to the analyzer and to the relays before replacing fuses.
Important Note: The analyzer circuit board components are sensitive to
static electricity damage in a powered or unpowered state. Do not touch
circuit boards or their components without wearing a properly grounded
wrist strap. See
Electrostatic Discharge (ESD) Considerations on page 111
DANGER: For continued
protection against fire hazard,
replace the fuse only with a fuse
of the same type and rating.
for
Spanish, French, German, or
Italian translations.
Figure 52 Fuse Locations
1.
PCB
2.
F3-Con25 & fan fuse: 1 A anti-surge 250 V (IEC127 Sheet III Type 2)
3.
F2-External valve fuse: 1 A anti-surge 250 V (IEC127 Sheet III Type 2)
4.
F5-Power supply terminal fuse: 2 A anti-surge 250 V (IEC127 Sheet III Type 2)
5.
Perforated power supply fuse: 4 A quick blow 250 V (IEC127 Sheet II Type 2)
6.
F4-RS422/485 fuse: 4 A anti-surge 250 V (IEC127 Sheet III Type 2)
7.
F1-UV lamp fuse: 4 A anti-surge 250 V (IEC127 Sheet III Type 2)
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...